• Top
  • 8615956659353
  • WhatsApp
Home    Lab instrments    Spectrometer    AAS    BK-AA4530F Atomic Absorption Spectrophotometer
白底阿里1000

BK-AA4530F Atomic Absorption Spectrophotometer

BK-AA4530F Atomic Absorption Spectrophotometer

BK-AA4530F Atomic Absorption Spectrophotometer

CHARACTERISTICS
1. Completely controlled by PC, can flexibly select the flame and graphite furnace atomizer (optional).
2. Integrated floated optical platform design can obviously improve the optical system shock resistance and keep stable though use the optical signal for a long term.
3. Eight light stands can be changed automatically and preheat the eight lights meantime as well as optimize the working condition of the hollow cathode lamp.
4. Position adjusting: the best height of the flame burner and can automatically set the front and rear positions.
5. Fully automated wavelength scanning and peak searching.
6. Complete safety chains protection equipment: the function of warning and automatic safety protection towards the wrong burner, leakage of the gas, under voltage of air and the abnormal flameout.
7. Deuterium lamp and self-absorption background regulation.
8. Data processing: super strong database, possesses more than 500 data self-storage and cut-off storage function, can store the analyzed result with the form of EXCEL and the testing method and the result can be randomly called.
9. Measuring method: flame absorption method and emission method.
10.Result printing: parameter printing, data result printing and diagram printing.

COMPLETE AUTOMATIC CONTROLLING SYSTEM
With the help of the software, the following can be easily achieved
1. Selection of the element lamp
2. Up-down-front-rear adjustment of the lifter
3. Adjustment of the optical energy
4. Selection of the slit
5. Determination of wavelength scanning and peak searching
6. Selection of the atomizer
7. Setting of the background deduction method
8. Controlling of the gas flow
9. Automatic flaming and flaming out
10.Setting of the graphite furnace testing method

ADVANCED GRAPHITE FURNACE TEMPERATURE CONTROLLING TECHNOLOGY
The bringing in of PID technology can effectively overcome the influence on the temperature rising process caused by the voltage fluctuation and the resistance change to make more accurate controlling process. The combination of the 3ms/time fast sampling technique can make more accurate and reliable testing data
The fast heating capacity can improve the flexibility of the elements further
Use the ordinary power source of 220V without need of dynamic power of 380V
The maximum procedure heating capacity setting of 20 levels can make a more convenient and easier test of different samples
Three grades adjustable gas flow can accustom to more application needs
Can timely alarm when the gas and water is stopped and insufficient gas and water, can avoid the equipment damage and measuring error

SAFE, RELIABLE AND CONVENIENT FLAME SYSTEM
EPC can control the flow of Acetylene (C2H2) more accurately and it is a kind of system which also can operate easily
Efficient a tomi zati on s ys tem enables a higher 

sensitivity
The whole operation system has a high security as the fire safety system can alarm whenever the electrici- ty is cut off, abnormal flame occurs, a lack of pressure happens or the burner does not match well. And it will automatically turn off the gas, prohibit temper ing. Thus it keeps the operation people and equipment from harm and damage.

MULTI-FUNCTIONAL SOFTWARE WORKSTATION

A work station that is supported WIND OW S XP/20 00/NT
The rich menu brings great convenience to the use of customer
Convenient conversion between different menus makes the operation more easily
Various analytical correction methods provide the users more choices
Basic default parameter settings enable even the beginners can do the normal operation
Flexible sto r age, e ditin g and p rin tin g m e th ods give the user largest suppor

AA4530F Atomic Absorption Spectrophotometer SPECIFICATIONS

Wavelength range 190 – 900nm
Spectral bandwidth 0.1, 0.2, 0.4, 1.0 and 2.0nm
Accuracy of the wavelength ± 0.15nm
Receptivity of the wavelength ≤0.04nm
Base line stability ≤0.002A/30min (Cu)
Characteristic viscosity 0.02μg/ml/1% (Cu)
Checking limit 0.004μg/ml (Cu)
Preciseness 0.5%
Grating 1800 lines
Inflamer All-metal titanium burner
Atomizer Effective glass atomizer
Lamp stand 8
D2 background correction ability When the background is 1 A, the background ability should be deducted not less than 50 times; 
self absorption background deduction method
Size and weight 700mm×550mm×450mm   75Kg
Power source 220V±22V AC

FLAME SYSTEM

Acetylene air burner 100mm
Ignition dynamic baseline drift   ≤0.006A/30min
(Cu) Characteristic viscosity ≤0.025μg/ml/1%
Related standard deviation of the accuracy ≤0.5%(Cu, absorbance>0.8A) (detection limit Cu≤0.008μg/ml)

Safety system  Can automatically cut off the gas when the pressure is not enough, the power is off, flameout and unconformity of the burner.

GRAPHITE FURNACE

The highest temperature  3000℃
The largest temperature rising speed ≥2000℃/S
Characteristic quantity Cd≤0.5×10-12g  Cu≤0.5×10-11g
Accuracy Cu≤3% Cd≤3%
Size and weight 550mm×450mm×300mm  65Kg
Safety system

Over current protection Low air pressure alarm / protection

Low cooling water flow alarm/protection

Power source and power 220V±22V  AC  7000W

STANDARD ACCESSORIES

PC workstation

HP inkjet printer

Oil free Air compressor

Acetylene reducing valve

Cu Hollow cathode lamp

Air filter