BK-GI T SERIES LCD Display Autoclave
BK-GI T SERIES LCD Display Autoclave
Introduction
“SMART IIH" microcomputer control system is adopted to realized not only intelligent controls of the sterilization process but also the classified management of instruments (user, administrator and engineer three levels included).As a result, we can guarantee satisfactory use and management of autoclaves.
Our patented pressure interlocking device can automatically sense the pressure in the sterilization chamber and tightly locked, which effectively prevent personal injuries caused by spraying steam arising from misoperation.The system can monitor the pressure changes in the sterilization chamber on a real-time basis. Upon detection of any pressure anomaly, the dual overpressure protection system consisting of a safety valve and a pressure protective device will be actuated immediately to relieve pressure or disconnect power supply.
The cover and bench of the sterilization chamber are made of special plastic to give a nice look and avoid causing corrosions and scald. The unique steam discharge and collection system can help to prevent scald.
5 sterilization cycles are preset for sterilization, drying, warming of solid and liquid articles and for melting and warming of agar separately.
Festures
Temperature and pressure
Sterilization temperature: 105 to 135°C,designed pressure:。.3Mpa
Big LCD Screen can clearly show all sterilization data and process.
Over temperature protection system
The system can monitor the temperature changes in the sterilization chamber on a real-time basis and disconnect power supply immediately upon detection of over temperature or abnormal temperature variation.
Automatic troubleshooting system
Troubleshooting is conducted automatically upon startup. Acoustic alarms will be given and Error Number will be displayed on the screen if any trouble occurs during equipment operation.
Safety and Function
Self-induction pressure interlocking device, lid closing checking, over temperature protection, temperature monitor, dry scorch protection system, over pressure protection, safety valve, over current and short circuit protection, leakage protection device, anti-scald safety protection, automatic troubleshooting system.
Type T: Semi-automatic, cooling fan, whole process shown by indicator
Type TW: Fully-automatic,standard+cooling fan, whole process shown by flow chart
Type TR: Fully-automatic, standard+cooling fan+drying function
FunctionTable
|
GI36TR/GI54TR |
GI36TW/GI54TW |
GI36T/GI54T |
|
GI80TR/GI100TR |
GI80TW/GI100TW |
GI80T/GI100T |
||
Sterilization mode |
Solid mode: Heating-Sterilizating-Exhausting |
Yes |
Yes |
Exhauting controlled by a knob |
Solidwithdrying mode:Heating-Sterilization-Exhausting-Water drainage-Drying |
Yes |
No |
No |
|
Liquid mode:Heating-Sterilizing-Precooling-Exhausing |
Yes |
Yes |
Exhauting controlled by a knob |
|
Liquid with warmingmode:Heating-Sterilizing-Precooling-Exhausting-Warming |
Yes |
Yes |
No |
|
Agar mode:Heating-Melting-Warming |
Yes |
Yes |
Yes |
|
Preset range of exhausting temperature |
73-104℃ |
73-104℃ |
Exhauting controlled by a knob |
|
Preset range of drying time&temperature |
1-300min & 80-140℃ |
No |
No |
|
Status flow chart that displays work processes |
Yes |
Yes |
Status indicator |
|
Timed startup |
0min-6days delayed |
0min-6days delayed |
No |
|
Procedures created |
60 |
60 |
20 |
Specification Table
Model |
GI36TW |
GI54TW |
GI80TW |
GI100TW |
GI36TR |
GI54TR |
GI80TR |
GI100TR |
|
Model difference |
TW: Standard + cooling fan, TR: TW+drying |
|||
Capacity(L) |
36 |
54 |
80 |
100 |
Dimension(L* W*H mm) |
460x542x1070 |
460x542x1070 |
620x680x1100 |
620x680x1200 |
Chamber dimension(Dia*H,mm) |
ф 325x460 |
ф325x670 |
ф 400x709 |
ф400x809 |
Chamber material |
SUS 304 |
|||
Rate power supply(w) |
2300 |
2600 |
4600 |
4600 |
Power requirement |
22OV,16A.50/60HZ 220V,32A. 50/60 HZ |
|||
Working environment |
5 45"C . relatively humidity 10%-85% |
|||
Stainless steel basket (Dia*H/mm) |
(ф 300x200)x2 |
(ф 300x200)x3 |
(ф360x280)x2 |
(ф360x220)x3 |
Temperature accuracy |
O.l°C |
|||
Working environment |
5 45"C, relatively humidity 10% 85% |
|||
Display system |
LCD displayer |
|||
Sterilization temperature |
105-135°C |
|||
Sterilization time range |
1-6000 minutes |
|||
Melting temperature |
60-100 |
|||
Melting time |
1-6000 minutes |
|||
Warming temperature |
45-60 °C |
|||
Warming time range |
1 9999 minutes |
|||
Drying temperature(TR) |
Drying time range 1-300 minutes, drying temperature: 80-140°C |
|||
Sterilization cycle |
|
|||
Sterilization cycle number |
Can save up to 60 cycles, record 1000 process, and USB port |
|||
Exhaust |
Automatic internal discharging , with a steam collecting bottle inside. Exhaust speed can be selected by 6 levels. bottle inside. Exhaust |
|||
Preset range of timer |
Can be preset and auto start up to 6 days in advance |
|||
Display range of pressure gauge |
-0.l-0.5Mpa |
|||
Pressure |
Max working pressure: 0.25Mpa; Design pressure:0.3Mpa |
|||
Controller |
"Smart IIII" fast speed microcomputer controller |
|||
Safety device |
Self-induction pressure interlocking device, lid closing checking, over temperature protection, temperature monitor; dry scorch protection system, over pressure protection,safety valve,over current and short circu it protection, leakage protection device,anti-scald safety protection, automatic troubleshooting system |